PENGARUH MEDIA KONTRAS TERHADAP NILAI HU, CTDI DAN DLP PADA PEMERIKSAAN CT SCAN UNTUK PENENTUAN NILAI DOSIS TIPIKAL = THE EFFECT OF CONTRAST MEDIA ON HU, CTDI AND DLP VALUES AT CT-SCAN EXAMINATION TO DETERMINE TYPICAL DOSE VALUES


SARI, NURUL AULIAH (2025) PENGARUH MEDIA KONTRAS TERHADAP NILAI HU, CTDI DAN DLP PADA PEMERIKSAAN CT SCAN UNTUK PENENTUAN NILAI DOSIS TIPIKAL = THE EFFECT OF CONTRAST MEDIA ON HU, CTDI AND DLP VALUES AT CT-SCAN EXAMINATION TO DETERMINE TYPICAL DOSE VALUES. Skripsi thesis, Universitas Hasanuddin.

[thumbnail of Cover]
Preview
Image (Cover)
H021211055-Cover.jpeg

Download (110kB) | Preview
[thumbnail of Bab1-2] Text (Bab1-2)
H021211055-1-2(FILEminimizer).pdf

Download (197kB)
[thumbnail of Dapus] Text (Dapus)
H021211055-dp(FILEminimizer).pdf

Download (79kB)
[thumbnail of Full Text] Text (Full Text)
H021211055-fulll(FILEminimizer).pdf
Restricted to Repository staff only until 18 June 2027.

Download (1MB)

Abstract (Abstrak)

Background. The increasing use of CT-Scan in medical diagnostics raises concerns about the radiation dose received by patients, especially when used with contrast media. This study aims to determine the effect of contrast media on Hounsfield Unit (HU), Computed Tomography Dose Index (CTDI), and Dose Length Product (DLP) values to support the determination of Typical Dose Values (NDT). Methods. This study uses an experimental quantitative approach by comparing CT-Scan examination data before and after contrast media administration in a number of patients. HU values are measured in the main blood vessels, while CTDI and DLP are obtained directly from the CT Scanner console. The study subjects consisted of 30 abdominal patients and 26 head patients for NDT analysis, and 20 patients for HU change analysis. Result. The results showed that there was a significant increase in HU values after contrast medium administration. For abdominal patients, the highest HU value from the three measurement points increased from 43.53 to 290.58. In the Head examination, the highest HU value increased from 35.98 to 67.65 after contrast medium was administered. Data analysis showed an increase in NDT after contrast use. The CTDIvol and DLP values were 5.05 mGy and 118.13 mGy.cm (abdominal CT), and 49.35 mGy and 1108.30 mGy.cm (head CT), respectively. Validation with DRL limit values was still below the standard. This study emphasizes the importance of optimizing the use of contrast media to obtain quality diagnostic images without ignoring radiation safety aspects. The conclusion of this study is that contrast media not only improves image quality by increasing HU values, but also affects the radiation dose received, so careful scanning parameter settings are needed to maintain a balance between image quality and radiation dose.

Keyword: contrast media, hounsfield unit, typical dose value, diagnostic reference level, abdomen, head.

Item Type: Thesis (Skripsi)
Uncontrolled Keywords: Contrast media, hounsfield unit, typical dose value, diagnostic reference level, abdomen, head.
Subjects: Q Science > QC Physics
Divisions (Program Studi): Fakultas Matematika dan Ilmu Peng. Alam > Fisika
Depositing User: Rasman
Date Deposited: 31 Oct 2025 07:31
Last Modified: 31 Oct 2025 07:31
URI: http://repository.unhas.ac.id:443/id/eprint/50400

Actions (login required)

View Item
View Item